-- DavidKeating - 20 Oct 2008

Module 15 - recovery after bleaching

The next draft will be compiled by (Messias and Jägle)

1. Scope

For dark-adapted conditions for visual cycle assessment

2. Identification

2.1. Number

15

2.2. Version number

1.0

2.3. Date of approval

3. Protocol use(s)

Assessment of rod phototransduction dynamics;

4. Patient population

Vitamin A deficiency;

RPE alterations;

Visual cycle alterations

5. Technical issues

4.1. Equipment and facilities

Step sequence must be automatic to ensure precise time intervals.

4.2. Staff expertise and responsibilities

No specific definition, see ISCEV Standard (2008 update) and Protocols general aspects.

4.3. Procedure

Pupil size recording

4.4. Patient Information

Bleaching

6. Response evaluation (with example)

Model to calculate recovery time rate.

7. Reporting and storage

Curves and amplitudes x time relationship.

8. Calibration

No specific definition, see ISCEV Standard (2008 update) and Protocols general aspects. 9. Protocol Specifications

Description

-1.0 Log cd.s/m²

[Pre-bleaching]

600 cd/m²

[Bleaching]

-1.0 Log cd.s/m²

[Post-Bleaching]

Averages per step

1

1

16

Time between averages

0

0

60 (s)

Adaptation time

0

240 (s)

0

Acquisition

Sample Frequency

1000 (Hz)

1000 (Hz)

1000 (Hz)

Sweep pre-trigger time

20 (ms)

20 (ms)

20 (ms)

Sweep post-trigger time

200 (ms)

200 (ms)

200 (ms)

Sweeps per result

5

1

5

Interstimulus interval

15000 (ms)

0 (ms)

15000 (ms)

Drift removal

Off

Off

Off

Manual rejection of sweeps

Off

Off

Off

DC off-set removal

On: 20 (ms)

On: 20 (ms)

On: 20 (ms)

Stimulus

Pulse frequency

Pulse duration (d)

4 ms

4 ms

Stimulus luminance

0.10 ph cd.s/m²

0

0.10 ph cd.s/m²

Stimulus color

White 6500K

White 6500K

White 6500K

Background luminance

0 cd/m2

600 ph cd/m2

0 cd/m2

Background color

White 6500K

White 6500K

White 6500K

Channels (1 and 2)

Filter low frequency cutoff

0.3 (Hz)

0.3 (Hz)

0.3 (Hz)

Filter high frequency cutoff

300 (Hz)

300 (Hz)

300 (Hz)

Y-axis scaling

50 (uV)

50 (uV)

50 (uV)

10. Relevant References

-- TWikiAdminUser - 11 Feb 2009

Comments on Recovery after bleaching

Back to Recovery after Bleaching

I have tried this protocol and have 2 problems with it:

1. The bleach is difficult for healthy subjects to comply with and IMPOSSIBLE for patients with central scotoma's to comply with. Have smaller bleaches been tested? Could Flash bleaches be used? This bleach cannot be used longitudinally, as no patient will consent to have this done to them twice.

2. You cannot average 5 stimuli with an ISI of 15 s each minute since the last flash of t1 must occur at the same time as the first flash of t2. I would suggest only averaging 3 flashes.

-Mitch

-- MitchellBrigell - 30 Mar 2009

 
Topic revision: r5 - 12 Feb 2009 - 18:57:11 - TWikiAdminUser
 
This site is powered by the TWiki collaboration platformCopyright &© by the contributing authors. All material on this collaboration platform is the property of the contributing authors.